Electrical Measuring Instruments Notes
ELECTRICAL MEASURING INSTRUMENTS
CHAPTER – 15 ELECTRICAL MEASURING INSTRUMENTS
THE MOVING COIL GALVANOMETER
The moving coil galvanometer is a basic electrical instrument. It is used
for the detection or measurement of small currents.
Principle
When current flows in a rectangular coil placed in a magnetic field it
experience a magnetic torque due to which it rotates through an angle
proportional to the current flowing through it.
Construction:
The essential parts of a moving coil galvanometer are
1. A U-shaped permanent magnet with cylindrical concave pole-pieces.
2. A flat coil of thin enamel Insulated wire (usually rectangular)
3. A soft iron cylinder
4. A scalar lamp and scale arrangement.
In suspended type or D Arsonvals galvanometer the flat rectangular coil of
thin enamel insulated wire of suitable number of terms wound on a light
non-metallic or non-magnetic (brass or aluminum) frame is suspected between the
cylindrical concave pole pieces of the permanent U-shaped magnet by a thin
phosphor-bronze strip. One end of the wire of the soil is soldered to strip.
The other end of the strip is fixed to the frame of the galvanometer and
connected to an external terminal. It serves as one current lead. The other end
of the wire of the coil is soldered to a loose and soft spiral of wire
connected to another external terminal. The soft spiral of wire serves as the
other current lead. A soft iron cylinder, coaxial with the pole-pieces is
placed within the frame of the coil but quite detached from it and is fixed to the
body of the galvanometer. In the space between it and the pole pieces, where
the coil moves freely, the soft iron cylinder makes the field stronger and
radial so that the magnetic field is always parallel to the plane of the coil.
To note the deflection a concave mirror along with lamp and scale arrangement
is used.
Working
When a current passes through the galvanometer coil, it experiences a magnetic
deflecting torque, which tends to rotate it from its rest position. As the coil
rotates it produces a twist in the suspension strip. The coil rotate until the
elastic restoring torque due to which the strip does not equalize and cancel
the deflecting magnetic torque and then it attains equilibrium and stops
rotating further.
i.e. Deflecting torque = Restoring torque
But deflecting torque = BINA Cos α
BINA Cos α = c0
Where B = strength of the magnetic field
I = current in the coil
A = Area of the coil
N = Number of turns in the coil
θ = Angle of twist of the suspension strip
c = torque per unit twist of the suspension strip for the equilibrium
As c/BNA is constant.
In pivoted type or Weston galvanometer the coil instead or being suspended by a
strip is pivoted between two jeweled bearings. The restoring torque is provided
two hair springs one on either side of the coil and curling on the opposite
sense are connected one to each spring. The hairspring thus also serve as
current leads to the coil. A light aluminum pointer is fixed to the coil, which
moves over a calibrated circular scale with equal divisions, which measures the
deflection directly.
Current Sensitivity of a Galvanometer
A galvanometer is said to be sensitive if for a small current the deflection
“θ” is sufficiently large. A galvanometer can be made more sensitive if c/BNA
is made small. Thus to increase sensitivity “c” may be decreased or B, N and A
may be increased “C” can be decreased by increasing the length of suspension
wire or by decreasing its can be decreased by increasing the length of
suspension wire or by decreasing its diameter, but this process cannot be taken
too far < as the suspension must be strong enough to carry the coil. “N” or
“A” cannot be increased because it makes the coil heavy. The loss of
sensitivity due to the use of fewer turns is however made us by the very high
value of the magnetic field employed.
The current sensitivity of a galvanometer is usually defined as the current in
microampere required to cause one-millimeter deflection on a scale place 1.0
meter from the mirror of the galvanometer.
THE AMMETER
Ammeter is an instrument, which is used for measuring electric current. A
galvanometer can measure small current if its scale is calibrated for the
current. For the measurement of large current a bypass resistance called a
shunt, of appropriate small value is connected in parallel with the
galvanometer coil. This resistance allows the large excess current through
itself while a fraction of the current passes through the galvanometer coil.
The scale of the instrument is so calibrated that it can measure the main
current directly.
Consider a galvanometer “G” whose resistance is “Rg” and which gives full scale
deflection when current “Ig” flows through it to convert the galvanometer into
an ammeter which can measure a maximum current “I” a shunt “Rs” of appropriate
small resistance should be connected in parallel with the galvanometer such
that the current “Ig” must flow through the galvanometer coil.
The potential difference “Vg” across the galvanometer is given by
Vg = Ig Rg
The potential difference “Vs” across the shunt is given by
Vs = Is Rs = (I – Ig) Rs
Where Is = I – Ig = current through the shunt.
As “Rg” and “Rs” are connected in parallel to each other therefore potential
difference across them will be equal i.e.
Vs = Vg
(I – Ig) Rs = Ig Rg
Rs = Ig Rg / I – Ig
Ammeter is always connected in series with the circuit.
Multi Range Ammeter
Sometimes an ammeter has more than one range, which means that it has as many
different shunts as the ranges. The desired range is selected by insertion the
proper shunt in position. In one type, one end of each shunt is permanently
connected to a common terminal while the other end of each is connected through
a range to a second common terminal.
THE VOLTMETER
Voltmeter is an instrument which is used for measuring potential difference
between any two points of a current carrying (or between the two terminals of a
source of emf). A galvanometer can be used for measuring a very small Potential
Difference. If its scale is calibrated for voltage. For the measurement of
large potential difference. A high resistance of the order of Kilo-ohms is
connected in series with it. This resistance is commonly known as “Multiplier
Resistance”.
Consider a galvanometer “G” where resistance is “Rg” and which deflects full
scale for the current “Ig” to convert this galvanometer into a voltmeter
measuring a Potential difference upto “V” volts. An appropriate high resistance
“Rn” must be connected in series with it such that for the potential difference
“V” applied between the ends of the above combination. The current “Ig” must
flow through the galvanometer. However the total resistance between the
terminal a & b is Rn + Rg.
(Rn + Rg+ Ig = V
Rn + Rg = V / Ig
Rn = V / Ig – Rg
Voltmeter is always connected in parallel with the circuit.
Multirange Voltmeter
Sometimes a voltmeter has more than one range, which means it has as many
different resistance as the ranges. The desired range is selected by inserting
the proper resistance in position.
We have a common terminal marked (+) and as many other terminals as the ranges.
In the other type one terminal is common marked (+) while the different range
terminals can be connected by a range switch to the other common terminals.
WHEAT STONE BRIDGE
If four resistances R1, R2, R3 and R4 are connected end to end in order to
form a closed mesh ABCDA and between one pair of opposite junctions. A and C
cell is connected through a key K1 while between the pair of opposite junctions
B and D a sensitive galvanometer “G” is connected through another key K2. The
circuit so formed is called a “Wheatstone-bridge”.
In the above bridge if the key is closed first, some current flows through the
cell and the resistance R1, R2, R3 and R4. If the key K2 is also closed the
current will usually be found to flow through the galvanometer indicated by its
deflection. However if the resistance R1, R2, R3 and R4 (or at least one of
them) are adjusted, a condition can always be attached in which the
galvanometer show no deflection at all i.e. no current passes through it. Then
the potential difference between B & D must be zero i.e. B & D must be
at the same potential. This implies that
P.d. between A and B = p.d. between A and D
OR
V(AB) = V(AD)
P.d. between B and C = p.d. between D and C
OR
V(bt) = V(Dt)
Since no current flows through the galvanometer the current R1 equals that in
R2, say II and the current in R3 equals that in R4 say 12
As,
V(AB) = V(AD)
1(1) R(1) = 1(2) R(3) ——– (I)
Also,
V(BC) = V(DC)
1(1) R(2) = 1(2) R(4) ——— (II)
Dividing eq 1 by eq 2
1(1) R(1) / 1(1) R(1) = 1(1) R(1) / 1(1) R(1)
R(1) / R(1) = R(1) / R(1)
Under balanced condition if any three resistance are known then the fourth can
be found easily (i.e. wheatstone principle).
The wheatstone principle is used in Meter Bridge, the P.O. box Carey Foster’s
Bridge, Callender and Graffite’s Bridge etc.
METER BRIDGE
The Meter Bridge also called slide – Wire Bridge is an instrument based on
wheatstone principle. It consists of a long thick copper strip bent twice at
right angles. Two small portions are cut off from it near the bends to provide
the gaps across which two resistances are known one and an unknown may be
connected. Each of the three pieces of the strip is provided with binding
screws. A uniform wire (of magnetic or other) one meter long and of fairly high
resistance is stretched, along the side a meter scale is connected to the ends
of the strip.
For measuring an unknown resistance “X” it is connected in one gap of the Meter
Bridge and a standard resistance box “R” is connected in the other gap. A cell
and a galvanometer are connected. The jockey “J” is moved along the wire to
obtain the balance point D. Under balanced condition if the length of the wire
segment. A D toward X is Lx and the length of the wire segment CD towards R is
L(R) then their resistances are ρL(R) respectively.
Where ρ = resistance per unit length of the wire.
POST OFFICE BOX [P.O BOX]
Post Office Box is an instrument, which is based on wheatstone principle. It
was first introduced for finding resistance of telegraph wires and for fault –
findings work in the post and telegraph office that’s why it is called “Post
Office Box”. It is more compact and easier to use.
It consists of three sets of resistances P, Q and R. The arms P and Q called
the ratio arms, usually consists of three resistances each viz. 10, 100 and
1000 ohms so that any decimal ratio from 1:100 to 100:1 may be used. The third
arm “R” is an ordinary set of resistances. The unknown resistances “X” to be
measured forms the fourth arm. Introducing the ratios 1:1, 10:1, 100:1 in turn
the balance or null position is traced by adjusting “R”. Balance is usually
obtained at the ratio 100:1 for some value of “R”. With this value of “R” the
value of X can be easily be calculated using relation of Wheatstone bridge i.e.
P/Q = R/X
X = R Q/P
THE OHMMETER
The ohmmeter is a device used for the measurement of resistance. It consists
of a sensitive galvanometer “G”, adjustable resister “R” and a torch cell “E”
connected in series between two terminals A and B. The unknown resistance “X”
to be measured is connected between the terminals A and B. The resistance R is
so chosen that when the terminals A and B are short circuited (i.e. X = 0). The
galvanometer gives full-scale deflection when no connection is between A and B
(i.e. X = ∞). The galvanometer shows zero deflection for the value of X between
= and ∞. The deflection is small or large depending on the value of X. The
scale of the galvanometer is calibrated with different known values of X and
there the circuit serves as an ohmmeter to measure any unknown resistance
approx. The scale of the ohmmeter however is not linear.
Using different conditions of R is series and different shunts across the galvanometer
worked by range switches, the ohmmeter can be adopted for different accessories
for e.g. 1 Ω accuracy in tens of ohms, in hundreds of ohms, in thousands of
ohms, in mega ohms etc. Ohmmeter is not a very accurate instrument.
POTENTIOMETER
Potentiometer is device for measuring the p.d (voltage) between two points
of a circuit or the e.m.f of a current source. It consists of a uniform wire
stretched on a wooden board along a meter scale.
Consider a uniform resistance wire AB of length L and Resistance R, across
which is connected to a source of constant EMF (e.g. an accumulator) through a
key and a rheostat to adjust and maintain a constant current 1 through it.
As the current flows, the P.d. between A and B = V(AB) = IR
If one terminal of a wire is connected to A while other is moved on the wire AB
then instrument acts as a Potential Divider.
To find an unknown EMF of a cell or some other potential difference or the
ratio of the emf of two cells consider the circuit. The positive terminals of a
cell of unknown EMF “E(N)” and a standard cell of Emf E(N) are connected to the
terminal A. The negative terminals of both the cells are joint to the jockey
through a two way key and a sensitive galvanometer. Using the two-way key first
cell E(N) only is introduced into the galvanometer branch and balanced point C
and length L are found for it.
THE AVO-METER
An Avo-meter is an apparatus which is used to measure current, voltage and
resistance in other words it is an ampere, volts and ohms. It can measure direct
as well as alternative voltage and currents. It consists of a galvanometer with
different scales graduated in such a way that all the three quantities can be
measured. A selector-cum-range switch is provided. Its has its own battery. A
rectifier is also included in the instrument to convert A.C. into D.C. before
they pass through the Galvanometer.